Ordering Information →
Technical Item ↓ |
BGD 862 Xenon Tester Chamber |
Xenon Lamp |
6.5 KW water cooling long arc xenon lamp |
Light Filter |
Import originally from ATLAS, can simulate indoor or outdoor sunshine spectrum |
Exposure Area |
6,500 cm2 |
Wave Length of Xenon Spectrum |
Can select 300-800nm(simulate sunshine,filtrate all wavelength under 300nm)or280-800nm (Add UV wave,keep some short-wave of UV) |
Adjustable Irradiance |
0.35~1.5 W/ m2 (Single point of control: 340nm)
0.7~4.0 W/ m2 (Single point of control: 420nm)
30 W/m2~150 W/ m2 (Full spectrum :from 300~400nm) |
Lifetime of lamps |
2,000 hours |
Adjustable BPT |
RT+10℃~100℃ |
Temperature Stability |
±1℃ |
Temperature Uniformity |
≤2℃ |
Temperature Declination |
≤±1℃ |
Adjustable Humidity |
Light:10%-85% Dark:10%-95% |
Rotate speed of sample |
1~7 r/min(circle as lamp centre) |
Spray Time |
Can set spray continuous time and spray period |
Machine producing deionized water |
Size:620*650*1600(L*W*H)
Weight:120KG |
Water for Chamber |
High purity deionized water(Conductivity<2us/cm) |
Flow of Deionized Water |
Adding humidity:0.2L/min;Spray:0.2L/min |
Overall Size |
1,200*1,200*2,000mm(L*W*H) |
Weight |
700 KG |
Structure |
Cabinet Material |
Whole chamber is made up of good quality stainless steel (SUS304) |
Chamber Door |
Simple door to left with filter window;silicon rubber seal the door edge |
Sample Crack |
Stainless steel material, its position in the chamber can be adjusted
(up, middle or down), Sample crack rotate around the centre of lamp |
Controller |
Siemens programmable controller. LCD touch-screen |
Way of adding humidity |
Add humidity by ultrasonic humidifier |
Black panel thermometer |
Heat insulate type black standard thermometer |
Cooling System |
Cooled by compressor and forced-air blast system |
Security Protections And Error Protections |
Xenon Lamp Protect |
Protection to cooling water for xenon lamp shortage.
Protection to big irradiance deviation for a long time |
Water Shortage Protection |
With water shortage protecting equipment in the water bath. Once the water bath is out of water, the controller will alarm automatically and record it |
Power Source Protect |
Protection for shortage phase and phase sequence |
Working Room Temperature Protect |
Individual over-temperature protection of working room temperature.
Over-temperature protection for BPT |
Pure System Protect |
Monitor the conductivity of pure water at real-time
(Once the conductivity≥2us/cm, the machine will stop and alarm) |
Protection to Radiation |
Xenon lamp will turn off automatically when opening the chamber door, in case xenon lamp radiate operator |
Installing Environment Requirement |
Power Source |
380V土10% 3-phases 4 lines or 220V土10% 2-phase; 50 or 60Hz;Electric current≥60A |
Air Outlet |
There should be air outlet in the indoor for convenient to connect vent pipe |